An Explanation of Sensitivity and the LLD, LLOQ, and ULOQ of a Multiplex ELISA

Assay sensitivity refers to the ability of a method or instrument to detect an analyte at a specified concentration and is often defined by a detection limit. There are a number of different “detection limits” that are commonly used in scientific literature and government agencies, but even when the same terminology is used, there can be differences according to what type of noise contributes to the measurement, calibration methods, instrument precision, etc.

At Quansys Biosciences, we often use three terms to define the range and sensitivity of our multiplex ELISA assays: Lower Limit of Quantification (LLOQ), Lower Limit of Detection (LLD), and Upper Limit of Quantification. The definitions of these terms as used at Quansys Biosciences and in science are discussed below:

Upper and Lower Limit of Quantification (ULOQ and LLOQ)
The ULOQ and LLOQ are the highest and lowest standard curve points that can still be used for quantification; they are the values below and above which, respectively, quantitative results may be obtained with a specified degree of confidence, or the highest/lowest concentration of an analyte that can be accurately measured. Together, the ULOQ and LLOQ define the range of quantification for the assay. Limits of quantitation are matrix, method, and analyte-specific, and can be calculated as follows:

Equation 1. (Calculation used in Q-View): ULOQ & LLOQ = Highest or Lowest Standard, respectively, with a %backfit of 120%-80%, a %CV of < 30%, and a positive mean pixel intensity difference between it and the negative control. Equation 2. (Commonly used in science to estimate the LLOQ): LLOQ = (Mean negative control pixel intensity) + 10 * (StDev of negative control pixel intensities). Lower Limit of Detection (LLD) The LLD is the lowest concentration level that can be determined to be statistically different from a blank at a 99% confidence level. In other words, it is the lowest quantity of a substance that can be distinguished from the absence of that substance (a blank value) within a stated confidence limit, generally 1%. The Limit of detection is matrix, method, and analyte-specific, and can be calculated as follows: Equation 1. (Calculation used in Q-View): LLD = 2*(StDev of negative control pixel intensities before ‘Negative Well Subtraction’)*LLOQ/ (Difference between pixel intensity of lowest standard and negative control) Equation 2. (Commonly used in science to estimate the LLD, which estimate is typically deemed acceptable if it falls within a region where the signal to noise ratio is greater than 5): LLD = (Mean negative control pixel intensity) + 2 * (StDev of negative control pixel intensities). Not only are the ULOQ, LLOQ, and LLD valuable measures of range and sensitivity for an individual assay, they can also be used to compare one experimental run to another, or watched over time as a parameter of assay stability. Reporting When reporting assay data, it is conventional to include values such as the ULOQ, LLOQ, and LLD. Data that falls outside the limits should be reported as > ULOQ, < LLOQ, or < LLD. Data extrapolated beyond the limits are typically not included in published results.

As researchers and problem solvers ourselves, we understand the value of sound data. We are proud to be a part of research that can better the world. We value the relationships, partnerships, and friendships that we have built with the people who trust and use our technology. We are committed to building these relationships. You can count on us to answer the phone and take time to thoroughly address questions or concerns about any of our products. In an industry that is reputed for grandiose claims, we trust that our quality standards and our customer service set us apart from the competition. If you’re happy with our products or if you think we can do something better, we hope you will let us know.

1-888-782-6797

info@quansysbio.com

365 N 600 W, Logan, UT